Pressure Necrosis and Osseointegration

Bookmark This Page

Introduction

The purpose of this paper is to discuss the risks and benefits of high insertion torque and its perceived correlation to the phenomenon called osseous pressure necrosis. Current evidence has not shown that properly applied high insertion torque leads to pressure necrosis when using the BIOMET 3i Tapered Implant System.

Background

As implant dentistry continues to evolve, there is increasing interest in providing patients with immediate functional and aesthetic replacement for a missing tooth or teeth. In an effort to achieve this outcome, clinicians are faced with several unique biomechanical challenges. Traditional dental implant protocols require the clinician to place an implant and allowit to osseointegrate (biologically stabilize) before it is exposed to the forces of occlusion and parafunction.1-3

Newer protocols involve reduced healing periods, the use of shorter length implants, inserting implants into poorer quality bone, and placing a restoration on an implant immediately. Some clinicians are inserting implants within an extraction socket or sockets coupled with immediate restoration.4-7

These newer protocols expose the implant to potentialmechanical stresses prior to biologic integration and the stresses may place the implant at risk for micromovement early within the healing process. It is well documented that excessive micromovement is detrimental to osseointegration and may well be the most common cause of implant failure, especially when following an immediate placement and immediate restoration protocol.8-11

In an effort to avoid this common complication, researchers have suggested controlling micromovement by splinting several implants together and/or ensuring that the implant has high primary stability.12-16 A common technique for evaluating primary stability is to monitor insertion torque on the drill unit. It is generally accepted that implants placed with higher torque have better survival rates than those seated with torque in the 20Ncm range.17 Several of the authors have tested numerous drill units and have evaluated torque accuracy. Due to potential variability in drill unit torque accuracy, the authors recommend use of a calibrated hand wrench if one desires to accurately determine insertion and final seating torque.

Insertion torque may be defined as the amount of torque required to advance the implant into the prepared osteotomy. Final seating torque may be defined as the torque measurement recorded when the implant reaches its final apico-occlusal position. While there have been several attempts to evaluate primary stability, one available method involved quantifying insertion torque. Historically, clinicians have used an additional method of measuring implant stability by using an Osstell Mentor Device (Osstell AB, Göteborg, Sweden) to obtain an Implant Stability Quotient (ISQ) value.

The literature includes studies that describe standards for both insertion torque18, 19 (cutting torque) and ISQ values,20-22 which may correlate to higher primary stability. These values have been correlated with a lower potential risk of excessive micromovement and subsequent implant failure. Minimal insertion torque values have been suggested to b at approximately 35Ncm. While some researchers suggest an ISQ value of at least 60,23, 24 others have also suggested that implants with an ISQ over 70 may be indicated for an immediate-load protocol.25

Historically, some researchers have relied on ISQ values as a significant part of the data presented in their research. However, the authors as well as some current researchers feel that the lack of clearly defined reference values makes the clinical use of the Osstell Device alone unreliable.26 Therefore, relying upon ISQ readings in lieu of clinical interpretation of implant insertional torque, is not recommended.

Nonetheless, use of this device remains a suitable adjunct for confirming the extent of implant primary stability.Modified drilling guidelines and new implant macrogeometric designs have evolved. The driver of these new designs and protocols has been a desire to improve primary stability and to reduce the risk of excessive micromovement during the early healing process, which involves the transition from primary stability to osseointegration. It is commonly accepted that the osseointegration process takes several weeks to months.27, 28

While the optimal values for insertion torque and ISQ values remain unknown, there is a general consensus that achieving higher Initial Bone-to-Implant Contact (IBIC) throughout the length of the osteotomy may correlate with achieving higher ISQ values and higher insertion torque values.29-31

While there are no known papers evaluating and comparing insertion torque to final seating torque and relating the numbers to implant survival (particularly in early and immediate load situations), there is a general consensus among the authors of this paper, that low insertion torque torque is less desirable and may provide a false sense of security when compared to a higher insertion torque coupled with a high final seating torque. The latter appears to provide the more desirable finding of top-tobottom stability.
In the authors’ opinion, it is crucial that the reader understand the purpose of high insertion torque. High torque, per se, is not the objective, but rather high insertion torque is an indicator of three-dimensional primary stability. The implant surgeon must think not only in terms of mechanical force, but of the fact that the force reflects an intimate three-dimensional contact between the walls of the osteotomy and the surface of the implant. High insertion torque must not be confused with high final seating torque, which may not imply threedimensional top-to-bottom contact providing optimal primary stability throughout the length of the implant. Trisi et al reported when implants were placed with more than 100Ncm of torque, the implants were always below the risk threshold for micromovement, which may lead to implant failure.33 In a clinical study, Ottoni et al showed that increasing peak insertion torque may significantly improve the survival rates of immediately loaded implants.34

As stated, high insertion torque and high ISQ values may be considered “go ahead” indicators by some clinicians when following newer biomechanically challenging protocols.While preliminary reports of clinicians and researchers following these new guidelines appear to be producing positive outcomes, others are concerned that higher insertion torque may be responsible for a phenomenon called osseous pressure necrosis.35-37

The term osseous pressure necrosis, although used frequently, has never been clearly defined in the dental literature. The authors define osseous pressure necrosis as excessive compression (pressure) of bone created during implant insertion. Compression of bone beyond its physiologic limits may result in ischemia leading to osseous necrosis. This phenomenon is generally viewed as being limited to cortical bone.35, 36

Some feel this process may lead to early bone loss and implant failure. The purpose of this document is to support new clinical recommendations and enhanced implant designs along with their use in accelerated treatment protocols, while reducing concerns about pressure necrosis.

Pressure Necrosis and Osseointegration

If pressure necrosis exists, the most likely causes are from one of three potential sources: the macrogeometry of the implant, the macrogeometry of the osteotomy site and the bone quality. It is generally accepted that the phenomenon would only be present in cortical bone.35-37

Although the placement guidelines for the BIOMET 3i Tapered Implant involve high insertion torque, bone taps are a part of cortical bone. This tapping process reduces implant insertion torque in cortical bone. The authors feel that the major risk of high insertion torque may be deformation of the implant’s macrogeometric features (i.e. stripping of the internal hex or driver tip) or an inability to fully seat the implant into dense bone. Further, it is the opinion of the authors that it is more for this reason rather than concerns over pressure necrosis that dense bone guidelines are required for placement of this properly macrogeometrically designed implant.

The BIOMET 3i Enhanced Tapered Implant’s depth and diameter specific drills replicate the minor diameter of the implant. Additionally, the design of these drills includes a built-in countersink, which accommodates the collar dimension of tapered implants. Therefore, compression of the implant against the walls of the osteotomy is unlikely with this macrogeometric relationship. Under these conditions, insertion torque may be caused by the threads simply cutting their way into the adjacent host bone (cutting torque
resistance).38


Each implant system must be evaluated to make sure its unique macrogeometric relationships and dense bone protocols do not produce excessive compressive stresses on the associated osseous structures or place the implant at risk for macrogeometric failure.

The lead author has followed and compared initial insertion torque on numerous BIOMET 3i Tapered Implants placed at higher torques with an average of 90Ncm. ISQ values were recorded initially and on several implants at 4, 8 and 12 weeks. ISQ values remained stable throughout the test period within 5%. Additionally, no unusual radiographic or crestal bone changes were noted.

“Therefore, compression of the implant against
the walls of the osteotomy is unlikely with
this macrogeometric relationship. Under these
conditions, insertion torque may be caused by the
threads simply cutting their way into the adjacent
host bone (cutting torque resistance).3838

Khayat et al presented results with Zimmer Tapered Implants inserted with torques up to 176Ncm. The implants were followed for one year and no signs of pressure necrosis, crestal bone change or untoward healing were noted when compared with the control group.39
“Current histologic research performed by Trisi
evaluating the effect of high insertion torque
on osseointegration, failed to demonstrate any
adverse outcomes.
“38


Finally, current histologic research performed by Trisi evaluating the effect of high insertion torque (up to 150Ncm with a mean torque of 110Ncm) on osseointegration, failed to demonstrate any adverse outcomes. As a matter of fact, this 1-to-45 day study demonstrated benefits including accelerated rates of osseous remodeling without any adverse effects on implant stability as a result of placing implants at what is traditionally considered high insertion torque.40

“Current histologic research performed by Trisi
evaluating the effect of high insertion torque
on osseointegration, failed to demonstrate any
adverse outcomes.

Conclusion

In conclusion, while osseous pressure necrosis may potentially adversely impact osseointegration, it is the experience of the authors that implant failure due to high insertion torque has not been noted. Furthermore, there is no clinical or histologic evidence to support the hypothesis that high insertion torque applied to an appropriately designed implant system leads to osseous pressure necrosis and possible subsequent implant failure. To the contrary, with the placement of BIOMET 3i Tapered Implants at high insertion torque, osseous pressure necrosis has not been noted by the authors. Rather, the authors concluded that high torque may positively impact traditional implant protocols, as well as, immediate placement and immediate restoration protocols.

References
1. Brånemark PI, et al (eds.) Tissue integrated prostheses: Osseointegration in
clinical dentistry. Chicago. Quintessence. 11-61.
2. Brånemark PI, et al. Osseointegrated implants in the treatment of the
edentulous jaw. Experience from a 10-year period. Scand J Plast Reconstr Surg
1977; 16(Suppl):1-132.
3. Soballe, et al. HA coating modifies implant membrane formation. Controlled
micromotion studied in dogs. ACTA Orthop Scand 1992; 63(2):128-140.
4. Ericsson, et al. Early functional loading of Brånemark dental implants: A 5-year
follow-up study. Clin Impl Dent Rel Res 2000; 2:70-77.
5. Malo, et al. Immediate function of Brånemark implants in the esthetic zone:
retrospective clinical study with 6-month to 4-year follow up. Clin Dent Rel Res
2000; 2:138-146.
6. Wohrle P. Single tooth replacement within the esthetic zone. Pract Peri Esth
Dent 1998; 10:1107-1114.
7. Pieri F, et al. Immediate occlusal loading of immediately placed implants
supporting fixed restorations in completely edentulous arches: A 1-year
prospective pilot study. J Perio 2009; 80:411-421.
8. Brunski JB. Avoid pitfalls of overloading and micromotion of intraosseous
implants. Dent Implantol Update 1993; 4(10):77-81.
9. Cameron H, et al. The effect of micromovement on the bonding of porous metal
to bone. J Biomed Mater Res 1973; 7:301-307.
10. Pilliar, et al. Observations on the effect of movement on bone ingrowths into
porous surfaced implants. Clin Orthop Relat Res 1986; 208:108-113.
11. Szmukler-Moncler S, et al. Timing of loading and the effect of micromotion on
bone-implant interface; Review of experimental literature. J Biomed Matr Res
1998; 43(2):192-203.
12. Pieri F, et al. Immediate occlusal loading of immediately placed implants
supporting fixed restorations in completely edentulous arches: A 1-year
prospective pilot study. J Perio 2009; 80:411-421.
13. Anitua, et al. Clinical outcome of immediately loaded dental implants bioactivated
with plasma rich growth factors; A 5-year retrospective study. J Periodontal
2008; 79:1168-1176.
14. Bergkvist G, et al. Immediately loaded implants supporting fixed prostheses in
the edentulous maxilla: A preliminary clinical and radiographic report. Int J Oral
Maxillofac Impl 2005; 20:399-405.
15. Tarnow DP, et al. Immediate loading of threaded implants at Stage 1 surgery in
edentulous arches: 10 consecutive case reports with 1 to 5 year data. Int J Oral
Maxillofac Impl 1997; 12:319-324.
16. Ibanez JC, et al. Immediate occlusal loading of double acid-etched surfaced
17. Ottoni JM, et al. Correlation between placement torque and survival of single
tooth implants. Int J Oral Maxillofac Impl Sept-Oct 2005; 20(5):769-776.
18. O’Sullivan D, et al. A comparison of two methods of enhancing implant
primary stability. Clin Imp Dent Relat Res 2004; 6:48-57.
19. Friberg B, et al. A comparison between cutting torque and resonance frequency
measurements of maxillary implants: A 20-month clinical study. Int J Oral
Maxillofac Surg 1999; 28:297-303.
20. Meredith N. Assessment of implant stability as a prognostic detriment. Int Journ
Pros 1998; 11:491-501.
21. Meredith N, et al. Resonance frequency measurements of implants in vivo. A
cross sectional and longitudinal study of resonance frequency measurements in
the edentulous and partially dentate maxilla. Clin Oral Implants Research
1997;8:226-333.
22. Friberg B, et al. Stability measurements of one stage Brånemark implants during
healing in mandibles: A clinical resonance frequency analysis study. Int J Oral
Maxillofac Surgery 1999; 28:266-272.
23. Nedir R, et al. Predicting osseointegration by means of implant primary stability:
A resonance-frequency analysis study with delayed and immediately loaded ITI
SLA implants. Clin Oral Impl Res 2004; 15:520-528.
24. Atsumi M. Methods used to assess implant stability current status. Int J Oral
Maxillofac Impl 2007; 22:743-754.
25. Nedir R, et al. Predicting osseointegration by means of implant primary stability:
A resonance-frequency analysis study with delayed and immediately loaded ITI
SLA implants. Clin Oral Impl Res 2004; 15:520-528.
26. Hobkirk JA andWiskott HW. Biomechanical aspects of oral implants: Consensus
report of working group 1. Clin Oral Impl Res 2006; 17 Suppl 2:52-54.
27. Brånemark PI, et al. Osseointegrated implants in the treatment of the edentulous
jaw: Experience from a 10-year period. Scand J Plastic Reconstr Surg 1977;
16(Suppl):1-132.
28. Schroeder A, et al. The reactions of bone, connective tissue and epithelium to
endosteal implants with titanium plasma sprayed surfaces. J Maxillofac Surg
1981; 9:15-25.
29. Mombelli A, Lang N. Clinical parameters for the evaluation of dental implants.
Periodontology 2000 1994; 4:81-86.
30. Albrektsson T, Albrektsson B. Osseointegration of bone implants: A review of an
alternative mode of fixation. Acta Orthop Scand 1987; 58:567-577.
31. Bahleitner A, Monov G. Assessment of bone quality: Techniques, procedures
and limitations. In Watzek G(ed). Implants in qualitatively compromised bone.
Chicago: Quintessence 2004; 55-66.
32. Turkyilmaz I, et al. Two alternative surgical techniques for enhancing primary
implant stability in the posterior maxilla: A clinical study including bone density,
insertion torque and resonance frequency analysis data. Clin Implant Dent Relat
Res, Dec 2008; 10(4):231-237.
33. Trisi P, et al. Implant micromotion is related to peak insertion torque and bone
density. Clin. Oral Impl. Res 2009;20:467–471.
34. Ottoni JM, et al. Correlation between placement torque and survival of singletooth
implants. Int J Oral Maxillofac Impl 2005;20:769-776.
35. Winwood K, et al. The importance of elastic and plastic components of strain in
tensile and compressive fatigue of human cortical bone in relation to orthopedic
biomechanics. J Musculoskelet Neuronal Interact 2006; 6:134-141.
36. Haider R, et al. Histomorphometric analysis of bone healing after insertion of
IMZ-1 implants independent of bone structure and drilling method (in German).
Z Stomatol 1991; 88:507-521.
37. Bashutski JD, et al. Implant pressure necrosis: Current understanding and case
report. J Periodontal 2009; 80:700-704.
38. Johansson P, Strid K. Assessment of bone quality from cutting resistance during
implant surgery. J Oral Maxillofac Impl 1994; 9:279-288.
39. Khayat P, et al. Clinical outcome of tapered implants placed with high insertion
torques (up to 176 ncm): Short oral communication (oral presentation –
Abstract 59) Clinical Oral Implants Research, Sept 2008; 19(9):855.
40. Trisi P. Immediate or late reconstructions in partially edentulous patients: Effect
regarding hard tissue aspects. 15TH Annual Scientific Meeting of the European
Association of Osseointegration, Kongresshaus Zurich, Zurich Switzerland;
October 6, 2006.

Authors and Affiliations

Alan Meltzer, DMD, MScD
Associate Clinical Professor, Department of Postgraduate Continuing Education
New York University School of Dentistry; Private Practice, Voorhees, NJ.

Harold Baumgarten, DMD*
Clinical Professor, Department of Periodontics, University of Pennsylvania
School of Dental Medicine, Philadelphia, PA; Private Practice, Philadelphia, PA.

Tiziano Testori, MD, DDS¶
Head, Section of Implantology, Department of Odontology, Galeazzi Orthopedic Institute;
University of Milan, Italy; Private Practice, Como, Italy.